PRODUCT FEATURE
- Available with a built-in blower, or without one for use with a central pump.
- Supports instant and customizable detection tasks.
- Enables real-time data transmission to EMC System software or other existing management systems.
PERFORMANCE FEATURES
- Particle size range: 0.3μm - 25μm
- Flow rate: 28.3LPM ± 5%
- 316L stainless steel, resistant to commonly used disinfectants
- Long-life laser with temperature control, corrosion-resistance and a 10-year warranty
- Easily integrated with user-owned systems to form a comprehensive monitoring and management platform for facilities or buildings
- Alarm light indicators for laser error, fow rate error and preset count limits
ACCURATE & RELIABLE
- Elimination of all paper transcription errors
- Improvements in sample integrity and accuracy of data collection
- Local visual indications of operation, stop, hardware error and networking mode
- ISO 21501-4:2018 compliant
DATA & TRENDS
|
Specification |
Airborne Particle Counter-Remote | Remote APC |
|||
|
A530BS |
A530B |
A530 |
||
|
Flow Rate |
28.3LPM ± 5% |
|||
|
Size Range |
From 0.3µm to 25µm |
|||
|
Size Channel |
0.3µm, 0.5µm, 5.0µm |
|||
|
Size Resolution |
<15% @ 0.5µm (ISO 21501-4) |
|||
|
Count Efficiency |
50% ±20% at 0.3µm, 100% ±10% for particles >0.45µm (ISO 21501-4 and JIS B9921) |
|||
|
Concentration Limit |
450,000 particles/ft³ at 10% coincidence loss |
|||
|
Zero Count |
<1count/5min |
|||
|
Exhaust |
Internal HEPA filter (>99.999%@0.3µm) |
N/A |
||
|
Flow Rate Control |
Built-in mass flow sensor, closed-loop auto control. |
Fixed flow orifice |
||
|
Laser Source |
Long-life laser | |||
|
Communication |
RJ45, USB, SENSER-HUB, WIFI |
|||
|
Data Storage |
4GB, default cache for 15,000 records, expandable up to 5 million records |
|||
|
Alarm |
Audible built-in alarm, external sound and light alarm |
|||
|
Display |
4"LCD capacitive touch screen |
status indicators |
||
|
Language |
Chinese, English |
N/A |
||
|
Blower |
Built-in blower |
N/A |
||
|
Calibration Frequency |
Recommended once a year |
|||
|
Calibration Standard |
Meets ISO 21501-4 |
|||
|
Dimensions (H×W×D) |
163×153×125 mm/ |
162×148×110 mm/ |
182×144×90 mm/ |
|
|
Weight |
1.95 kg/ 4.30 lbs |
1.85 kg/ 4.08 lbs |
1.60 kg/ 3.53 lbs |
|
|
Enclosure |
Aluminum or 316L stainless stee |
316L stainless steel |
||
|
Power |
100-240V, 50Hz/60Hz AC to 24V DC input, plug compatible with standard interface types |
|||
|
Max Power Consumption |
AC 90W |
|||
|
Operating Conditions |
Operating temperature: 5℃-35℃ (41℉-95℉) |
|||
|
Storage Conditions |
Storage temperature: 0℃-40℃ (32℉-104℉) |
|||
|
Safety |
EN 61010-1:2010, EN 61326-1:2013, EN IEC 61326-1:2021 |
|||
|
Warranty |
24 months (calculated from the date of product activation or six months after the date of manufacture, whichever comes first). 10 years for laser. |
- User-friendly interface
- Streamlined sampling scheme modifcation for added fexibility
- Intuitive touch screen display
- Storage of up to 15,000 sample datasets
- Continuous real-time data collection to support trending
INSTALLATION MODE


Humidification
Dehumidification
HVAC Measuring instruments
HVAC Sensors & Field Devices
Air Particles Counting